Difference Detection Matrix Example: Difference between revisions

From ReliaWiki
Jump to navigation Jump to search
No edit summary
No edit summary
Line 10: Line 10:
[[Image: Detection Matrix Example Result.png|thumb|center|400px]]
[[Image: Detection Matrix Example Result.png|thumb|center|400px]]


From the  
In the above figure, the range of the assumed B10 life for each design is from 500 to 3,000. The pair-wised comparison results are displayed in each cell. 0 means the difference cannot be detected through the test; 1 means the difference can be detected if the test duration is 5,000 hours; 2 means the difference can be detected if the test duration is 3,000 hours. For example for cell (1000, 2000), the number is 2. It means if the B10 life for design 1 is 1000 hours and the B10 life for design 2 is 2000 hours, by testing 20 samples each for 3,000 hours, the difference of their B10 lives probably can be detected. This is because at a confidence level of 90%, their estimaed confidence intervals of the B10 life do not overlap. This can be seen in the figure below.


[[Image: Detection Matrix Example B10 Intervals.png|thumb|center|400px]]
[[Image: Detection Matrix Example B10 Intervals.png|thumb|center|400px]]


[[Image: Detection Matrix Example Parameter Estimator.png|thumb|center|400px]]
[[Image: Detection Matrix Example Parameter Estimator.png|thumb|center|400px]]

Revision as of 21:08, 24 February 2012

Test Design Using Life Difference Detection Matrix

Assume there are two design options for a new product. Engineers need to design a test to compare the reliability perfomance of these two options. Both designs are assumed following Weibull distribution. For Design 1, its shape parameter [math]\displaystyle{ \beta }[/math] = 3; for Design 2, its [math]\displaystyle{ \beta }[/math] = 2. Their B10 lives may range from 500 to 3,000 hours. We want to use the Difference Detection Matrix to choose the suitable sample size and test duration.


Solution

Initially, the sample size for each design is set to 20. Two test durations 3,000 and 5,000 also were chosen. Based on these settings and the assumptions on the failure time distributions, the following Difference Detection Matrix is obtained.

Detection Matrix Example Result.png

In the above figure, the range of the assumed B10 life for each design is from 500 to 3,000. The pair-wised comparison results are displayed in each cell. 0 means the difference cannot be detected through the test; 1 means the difference can be detected if the test duration is 5,000 hours; 2 means the difference can be detected if the test duration is 3,000 hours. For example for cell (1000, 2000), the number is 2. It means if the B10 life for design 1 is 1000 hours and the B10 life for design 2 is 2000 hours, by testing 20 samples each for 3,000 hours, the difference of their B10 lives probably can be detected. This is because at a confidence level of 90%, their estimaed confidence intervals of the B10 life do not overlap. This can be seen in the figure below.

Detection Matrix Example B10 Intervals.png


Detection Matrix Example Parameter Estimator.png
Detection Matrix Example Expected Failure Tiime.png
Detection Matrix Example Std Folio Data.png.png
Detection Matrix Example B10 QCP.png.png